HWS Horizontal Laminar Flow Cabinets

GET IN TOUCH

HWS Series cabinets are self-contained Class 3.5 horizontal laminar flow work stations designed for installation on standard laboratory benches, or on optional floor stands. Cabinets are produced with work zone width of 90cm, 120cm or 180cm, with the option of 60cm or 75cm work zone height.

APPLICATIONS

HWS Series Horizontal Laminar Flow Cabinets

HWS Series cabinets are designed to provide a high degree of protection for process products and apparatus in laboratory and production facilities. Many critical applications in the medical, pharmaceutical, scientific and electronics fields demand an ultra-clean work environment which is free from biological and particulate contamination.

Typical applications include:

  • Aseptic dispensing
  • Assembly of electronics and optics components
  • Cell culturing
  • Media pouring
  • Preparation of medical devices/equipment
  • Sterility testing

HWS Series cabinets are intended for work with non-hazardous materials in such applications.

Adheres to Australian and International Standards

DESCRIPTION

HWS Series cabinets are self-contained Class 3.5 horizontal laminar flow work stations designed for installation on standard laboratory benches, or on optional floor stands. Cabinets are produced with work zone width of 90cm, 120cm or 180cm, with the option of 60cm or 75cm work zone height.

Each cabinet is factory-tested and certified by a NATA-registered laboratory to establish compliance with the performance requirements of AS1386, Part 5.

Vertical Flow Cabinets
VWS Series cabinets may be preferred for applications where the work process is optimally presented to vertical airflow.

Operation
Horizontal laminar flow in the work zone creates a biologically-clean, particle-free work environment.

A direct-drive fan draws in ambient air through a prefilter on the top of the cabinet and supplies it to the work zone through a HEPA filter. The average air velocity in the work zone is maintained between 0.45 and 0.5 m/s, with all velocity readings within 20% of their average.

Air cleanliness within the work zone is better than Class 3.5 (Class 100). To enhance work zone air cleanliness, it is recommended that cabinets are installed in cleanrooms which provide a secondary barrier.

CONSTRUCTION

Cabinet
Constructed in electro-galvanised steel, etch primed and finished in a special baked enamel that has been developed for laboratory equipment. The work zone is constructed in Grade 304 stainless steel. The stainless steel work surface incorporates a lip to prevent liquid spills from contacting the HEPA filter. A removable, washable metal screen protects the HEPA filter form mechanical damage.

Fans
A direct-drive fan is regulated by a speed controller to enable airflow to be maintained through filter life. Fans and filter plenums are designed to provide ultra-quiet operation and low vibration level.

HEPA Filters
Clyde-Apac HEPA filters, are manufactured to the most exacting standards for critical applications. Each filter is individually certified to achieve an arrestance efficiency of not less than 99.995% to the stringent BS3928 Sodium Flame test.

All testing is conducted in a NATA-registered factory laboratory and a NATA-endorsed test label, being an extract of the test report, is affixed to each filter.

Prefilters
An easily-accessed, washable prefilter arrests not less than 90% of particles 5 micron and larger, thus prolonging HEPA filter life. A warning light in the control panel signals the need for Prefilter service.

PHYSICAL DATA

Model
Overall Dimensions
(mm)
Work Zone Dimensions
(mm)
Weight 
(kg)
W
D
H
W
D
H
HWS90
975
750
1130
880
500
585
115
HWS120
1285
750
1130
1185
500
585
128
HWS120/75
1285
750
1280
1185
500
735
145
HWS180
1895
750
1130
1795
500
585
180
HWS180/75
1895
750
1285
1795
500
735
198
HWS240
2515
750
1130
2415
500
585
232
HWS240/75
2515
750
1285
2415
500
735
250

Electrical
A low voltage touch control panel is located on the front of the cabinet. The status of all switched functions is indicated by LEDS. Glare-free lamps provide a minimum lighting intensity of 800 lux at the work floor.

  • Standard features
  • Stainless steel work zone
  • Low voltage touch controls
  • Prefilter service indicator
  • Glare-free fluorescent lighting
  • Quiet, low-vibration operation
  • Comprehensive operation manual

Options

  • Germicidal UV lamp
  • UV shield/ work zone cover
  • Hourmeter
  • Manometer
  • Power outlet
  • Service taps for air, gas and vacuum
  • Floor stand
  • UV sterilising chamber
  • Plastics work surface
  • ‘Metal-free’ construction
  • Heated work surface for IVF applications

Personnel Protection – Limitations
HWS cabinets provide protection for products or experiments, but do not protect personnel from aerosols of hazardous materials that may be handled in the cabinet.

For applications where personnel and environmental protection is required, Clyde-Apac Class I or Class II biological safety cabinets, or cytotoxic drug safety cabinets (as applicable) should be considered.

Quality Assurance and Control
Clyde-Apac laminar flow and HEPA filter products are manufactured in Australia under an accredited Total Quality Management system.

ON-SITE TESTING

HWS Series cabinets are factory-tested and certified by a NATA-registered laboratory. Additional testing and certification is recommended as follows:

  • On site prior to use
  • After any electrical or mechanical maintenance
  • After filter replacement
  • After re-location
  • At least annually
  • In special circumstances, e.g. if faulty cabinet operation is suspected

Laminar Air Flow Services, a NATA-registered laboratory and a division of Clyde-Apac, provides comprehensive on-site maintenance, testing and certification services for safety cabinets, laminar flow work stations, cleanrooms and HEPA filter installations. This service is available from fully-equipped, NATA registered laboratories in major Australian and selected International centres.

Due to ongoing research & development, Clyde-Apac reserves the right to modify the product & specifications without notice.

Clyde-Apac Laminar Air Flow equipment & HEPA filter products available from authorised dealers and distributors